Abstract


An amperometric flow system for glucose determination in blood serum samples after enzymatic reaction with glucose oxidase immobilized on magnetite covered with silica gel modified propylamine is described. The solid was magnetically retained on a mini-column and placed into the flow injection system preceding the amperometric detector using a modified screen printed electrode with [Fe(tris(3,5-dimetyl-1-pyrazolyl)borate)2]+[FeCl4]-. The variables involved in the system such as flow rate, enzyme concentration, injection volume and reaction coil length were evaluated using a Taguchi parameter design. Under optimal conditions, the calibration curve of glucose sample was linear between 0.24 and 6.00 mM, and with a limit of detection of 0.08 mM. The repeatability for a 4.0 mM glucose solution was 1.0%.The method was validated by comparing the obtained results to those provided by the enzymatic spectrophotometric method; no significant differences were observed.



Producto de Investigación




Artículos relacionados

Silver and Silver Chloride Electrodeposits, an Alternative in the Construction of Ag/AgCl Solid Elec...

Development of a novel nitrate-selective composite sensor based on doped polypyrrole

Synthesis, characterization, and crystal structures of n-alkyldiorganodithiophosphates RS2P(OC6H4)2

Determination of oxytetracycline in milk samples by polymer inclusion membrane separation coupled to...

Synthesis and crystal structures of cis-palladium(II) and cis-platinum(II) complexes containing dipy...

Potentiometric Behavior of Diverse Polypyrrole-sulphate Films Electro Synthesized on Graphite - Epox...

Potentiometric behavior of graphite-epoxy electrochemical transducers towards anions, cations and pH...

Prevalence and Antimicrobial Resistance Patterns of Salmonella from Different Raw Foods in Mexico

Validation of a tubular bismuth film amperometric detector: Determination of diclofenac sodium by mu...

Kinetics of Polypyrrole Films Doped with Sulphate Ions Electrodeposited Over Graphite - Epoxy Resin ...